Pavone UTOC1901 PS19 Off Center load cell

Pavone UTOC1901 PS19 Off Center load cell

Code: UTOC1901
Brand: Pavone
UTOC1901 PS19 30 KG
UTOC1902 PS19 75 KG


The off center load cell PS19 is made of stainless steel and has been designed for the construction of scales for electromedical equipment and for dosing systems. The single point cell PS19 can be customized: the load capacity can vary from 250 kg to 400 kg and the dimensions of the plate can vary from 600 x 600 mm to 800 x 800 mm for cells with a high load capacity.

Technical Specifications

Rated load RL: 15, 20, 30, 50, 75, 120, 200, 250, 350, 400 Kg
Combined error: < ±0.017 % RO
Repeatability: < ±0.010 % RO
Creep (30 minutes): ±0.016 % RO
Safe overload: 150 % RL
Ultimate overload: 200 % RL
Material: Stainless steel
Degree of protection: IP68
Accuracy class: 3000 OIML
Deflection: 0.3 ÷ 0.5 mm
Compensated Temperature: -10 ÷ +40 °C
Temperature range: -20 ÷ +70 °C
Temperature effect on zero balance: < ±0.01 % RO/5°K
Temperature effect on output: < ±0.006 % on output/5°K
Rated output RO: 2.0 ±0.2 mV/V
Zero balance: < ±2 % RO
Insulation resistance: > 5000 MOhm
Input resistance: 400 ±20 Ohm
Output resistance: 350 ±3 Ohm
Recommended input: 5 ÷ 15 Vcc/ca


Pavone Datasheets

Country & Language File Name Download Link
United States (English) Technical_Specifications_Pavone_Systems_dat400_ab4429_77453.pdf Download
United States (English) Technical_Specifications_Pavone_Systems_dat500_4f7480_64370.pdf Download
United States (English) Technical_Specifications_Pavone_Systems_dat-500-analog_efec3b_78468.pdf Download
United States (English) Technical_Specifications_Pavone_Systems_mc302_43103b_19246.pdf Download
United States (English) Technical_Specifications_Pavone_Systems_mct-1303-batch_081be9_91178.pdf Download
United States (English) Technical_Specifications_Pavone_Systems_tester1006_9b8a5a_88792.pdf Download

Request a Quote