The off center load cell PS19 is made of stainless steel and has been designed for the construction of scales for electromedical equipment and for dosing systems. The single point cell PS19 can be customized: the load capacity can vary from 250 kg to 400 kg and the dimensions of the plate can vary from 600 x 600 mm to 800 x 800 mm for cells with a high load capacity.

Technical Specifications

Rated load RL: 15, 20, 30, 50, 75, 120, 200, 250, 350, 400 Kg
Combined error: < ±0.017 % RO
Repeatability: < ±0.010 % RO
Creep (30 minutes): ±0.016 % RO
Safe overload: 150 % RL
Ultimate overload: 200 % RL
Material: Stainless steel
Degree of protection: IP68
Accuracy class: 3000 OIML
Deflection: 0.3 ÷ 0.5 mm
Compensated Temperature: -10 ÷ +40 °C
Temperature range: -20 ÷ +70 °C
Temperature effect on zero balance: < ±0.01 % RO/5°K
Temperature effect on output: < ±0.006 % on output/5°K
Rated output RO: 2.0 ±0.2 mV/V
Zero balance: < ±2 % RO
Insulation resistance: > 5000 MOhm
Input resistance: 400 ±20 Ohm
Output resistance: 350 ±3 Ohm
Recommended input: 5 ÷ 15 Vcc/ca


Pavone

United States (English) Technical_Specifications_Pavone_Systems_dat400_ab4429_77453.pdf
United States (English) Technical_Specifications_Pavone_Systems_dat500_4f7480_64370.pdf
United States (English) Technical_Specifications_Pavone_Systems_dat-500-analog_efec3b_78468.pdf
United States (English) Technical_Specifications_Pavone_Systems_mc302_43103b_19246.pdf
United States (English) Technical_Specifications_Pavone_Systems_mct-1303-batch_081be9_91178.pdf
United States (English) Technical_Specifications_Pavone_Systems_tester1006_9b8a5a_88792.pdf

Solicitar una Cotización